Monochromatic Calibration Light from 260–1500 nm
The LiSen Optics iSpec-ISP100-XE provides uniform illumination at selected wavelengths from 260 to 1500 nm. It combines a monochromator with an integrating sphere for camera calibration, spectral response measurements and other calibration work that requires a controlled wavelength and a uniform output.
The system includes a regulated DC power supply, an optical coupling module, an integrating sphere with an intensity monitor, a control unit and measurement software. The specified output uniformity is 98%. Factory calibration and live intensity readout let the operator monitor the sphere output during a measurement.

Wavelength and Output Control
Dedicated software sets the output wavelength, with step sizes of 1, 5 or 10 nm. Wavelength accuracy is ±0.2 nm and repeatability is ±0.1 nm. The high-power xenon source covers the ultraviolet portion of the specified range; the configuration table lists xenon and halogen sources.
Slit widths of 3, 2.7 and 1.4 mm are available. At the 1.4 mm setting, the specified spectral bandwidth is 10 nm full width at half maximum (FWHM), and integrating sphere output power is ≤300 μW. The power and bandwidth figures should be read with this slit condition.
A high-numerical-aperture optical design provides optical throughput greater than 90%. The regulated DC supply allows output intensity to be adjusted over a wide dynamic range. Source stability is specified as 0.1%.
Calibrated Output Monitoring
The integrating sphere includes an intensity detector and is supplied with factory output-intensity and spectral-irradiance calibration. The software displays the output level in real time, allowing the operator to monitor the source while setting the wavelength and intensity.
A regulated DC supply and optical breadboard are included. The system communicates through USB 2.0 or RS-232 and combines the light source, monochromator and sphere in a single setup.
Output Power and Spatial Uniformity
The output-power plot compares the 3 mm, 2.7 mm and 1.4 mm slit settings. The spatial uniformity figure shows the output distribution and the measurement arrangement used to assess uniformity across the sampled area.


Technical Specifications
| Parameter | iSpec-ISP100-XE |
|---|---|
| Product Image |
|
| Wavelength Range | 260–1500 nm |
| Source Stability | 0.1% |
| Slit Width | 3, 2.7, or 1.4 mm (optional) |
| Integrating Sphere Output Power | ≤300 μW at 1.4 mm slit width |
| Output Uniformity | 98% |
| Light Source | Xenon / halogen |
| Spectral Bandwidth (FWHM) | 10 nm at 1.4 mm slit width |
| Wavelength Step Size | 1, 5, or 10 nm |
| Wavelength Accuracy | ±0.2 nm |
| Wavelength Repeatability | ±0.1 nm |
| Integrating Sphere Size | 100 mm |
| Sphere Port Size | 30 mm |
| Intensity Monitor | Included |
| Regulated DC Power Supply | Included |
| Optical Breadboard | Included |
| Dimensions | 900 × 400 × 350 mm |
| Weight | 30 kg |
| Communication Interfaces | USB 2.0 / RS-232 |
| Optical Throughput (Optical Design) | >90% |
Calibration Applications
The iSpec-ISP100-XE is used for camera calibration, satellite remote sensing calibration measurements, and radiance and irradiance calibration. It also supports night-vision systems, security cameras, high-sensitivity imagers, and CMOS/CCD spectral response testing and calibration.
Explore the integrating sphere uniform light source range, or contact LiSen Optics to discuss the wavelength range, output level and calibration requirements for your instrument.